Overview

The use of fuel cell sensors to measure trace oxygen content is one of the current advanced measurement methods. It has the characteristics of fast measurement speed and high accuracy. Because the sensor is completely sealed, the sensor is maintenance-free, and the service life is usually different due to the frequency of analysis and the oxygen content of the sample. The portable trace oxygen analyzer developed and produced by our company uses precision resistors as sampling resistors to make the measurement accuracy higher. It uses high-precision operational amplifiers and low on-resistance, low-channel matching and low flatness multi-channel analog switches to achieve oxygen The automatic switching of the content range gear is an ideal instrument for portable trace oxygen analysis. It is widely used in petrochemical, steel, electric power, medical and other industries. Because the instrument is designed with a unique purging and sampling process, it plays an important role in fast and large workload analysis.

This instrument is suitable for the analysis of trace oxygen in non-acidic and non-corrosive gases.

trace Oxygen meter

Performance parameter

1. Measuring range: 0~21%;

2. Measurement accuracy: ±5%;

3. Response time: 90% of reading in 10 minutes;

4. Working temperature: -5℃~40℃;

5. Power supply: 8800mAh battery pack;

6. Working pressure: 0.04MPa;

7. Dimensions: length 300×width 250×height 150mm

8. The weight of the whole machine: about 10kg

Working principle

I. Working principle of fuel cell sensor

The fuel cell sensor is composed of a highly active oxygen electrode and a lead electrode, immersed in a KOH solution. Oxygen is reduced to hydroxide ions at the cathode, while lead is oxidized at the anode.

O2+2H2O+4e 4OH-

2Pb+4OH- 2PB(OH)2+4e

The KOH solution is separated from the outside by a polymer film, and the sample gas does not enter the sensor directly, so the solution and the lead electrode do not need to be cleaned or replaced regularly. The oxygen molecules in the sample gas diffuse through the polymer film into the oxygen electrode for electrochemical reaction. The current generated in the electrochemical reaction is determined by the number of oxygen molecules that diffuse to the oxygen electrode, and the diffusion rate of oxygen is proportional to the sample gas. Oxygen content. In this way, the sensor output signal size is only related to the oxygen content in the sample gas, and has nothing to do with the total amount of gas passing through the sensor. Through the connection of the external circuit, the charge transfer in the reaction, that is, the magnitude of the current, is proportional to the oxygen participating in the reaction.

Using this method to measure oxygen can not be affected by the reducing gas in the measured gas, eliminating many sample gas processing systems. It is faster than the old "Golden-Lead" primary battery to measure oxygen, and does not require a long boot-up blowing process. The sample gas of the "Golden-Lead" primary battery enters the solution directly, which leads to a large amount of maintenance of the instrument. In the fuel cell method, the sample gas does not enter the solution directly, and the sensor can work very stably and reliably for a long time. In fact, the fuel cell oxygen sensor is completely maintenance-free.

II. The working principle of the deoxygenation device, the stabilizing valve and the fuel cell sensor

The working principle between the deoxygenation device, the stabilizing valve and the fuel cell sensor is as follows:

1. The pressure stabilizing valve stabilizes the pressure of the sample gas at 0.04MPa, meeting the requirement that the fuel cell sensor's working pressure does not exceed 0.04MPa, and protects the fuel cell sensor from damage caused by human misoperation.

2. Deoxygenation tube: deoxygenate the sample gas and zero point correction

3. Plane four-way switching valve: switch between sampling/venting, deoxygenation purging/normal purging