Automatic Close Cup Flash Point Measuring Instrument I.Function Overview HZBS-X3 Close Cup Flash Point Tester is designed and made as per the ASTM-D93, it is an advanced analytical instrument, mainly used to determine the closed flash point of petroleum products. II.flash point measuring device price Technical Parameters 1....
Product Details ofHZBS-X3 Automatic Close Cup Flash Point Measuring Instrument

Huazheng Electric Manufacturing (Baoding) Co., Ltd. is one of the leading manufacturers and suppliers of hzbs-x3 automatic close cup flash point measuring instrument in China. Please feel free to buy high quality equipment for sale here from our factory. Good service and reasonable price are available.

 

HZBS-X3 Automatic Close Cup Flash Point Measuring Instrument

I.Function Overview

HZBS-X3 Close Cup Flash Point Measuring Instrument is designed and made as per the ASTM-D93, it is an advanced analytical instrument, mainly used to determine the closed flash point of petroleum products.

ASTM D93 Closed Cup Flash Point measuring equipment

II.Technical Parameters

1. Temperature Measurement

Ø Temperature measuring range: -59.9°C~299.9°C

Ø Temperature Control Range: Room temperature~300°C

Ø Repeatability: 0.025X(x=twice successive test average value)

Ø Resolution: 0.1°C

Ø Precision: 0.5%

Ø Temperature Sensor: Platinum Resistor(PT100)

Ø Flash Fire Sensor: Ion detection ring

2. Power Consumption: approx 400W

HZBSX3 flash point test equipment

III. Precautions and Instrument Maintenance

1) The instrument should be used in a non-corrosive environment.

2) The oil cup should be cleaned, when replacing test sample.

3) Check thermocouple should not have oil contamination, or it will affect the detection sensitivity.

4) No touch ignition head, oil cup and its surrounding parts to avoid scorching during the test.

IV.Failure and Treatment

NO.

Failure

Treatment

1

Turn on the power,the LCD screen no display

Check power supply is plugged in, open the instrument measured the board examine the plug is not loose.

2

Large repeatability error for sample testing

Remove the lifting arm top cover, and check

1. whether the mixer shaft is broken

2.Oil may effect on thermocouple sensitivity, plz dry it by filter paper

3

No heating

Electric wire broken

4

Lift arm rise or fall failure

Limit optocoupler damage

The screw between main engine and lift arm fell off

flash point test instrument

lQDPKcrtGtmqmW_NAazNAyCw_iF2foFjEn0JNdw6T7ADAg_800_428

Customer Visit

Case

Request for quote_

Hot Tags: HZBS-X3 Automatic Close Cup Flash Point Measuring Instrument, China, suppliers, manufacturers, price, for sale, oil flashpoint detector, Flash Point Test Machine, flash point measuring device, flash point measuring apparatus, Flash Point Measurement Equipment, flash point analyser HZBS-X3, automated pensky martens flash point analyzer, HZBSX3, automatic closed flash point measuring device, china flash point test equipment price, closed cup flash point measurement, open cup flash point meter, Cleveland Open Cup Flash Point Apparatus, open cup flash point test machine, manual flash point tester