HZJF-9041D Fourchannels Partial Discharge Test System

Oct 25, 2023

HZJF-9041D Fourchannels Partial Discharge Test System

 

Structure Description

The instrument is a desktop standard cabinet structure. The operation surface of the instrument is divided into two parts: front panel and back panel. The positions of each component are shown in Fig. 2a and 2b

 

news-967-555

Figure 2a HZJF-9041DSchematic diagram of front panel of partial discharge detector

news-899-539

Figure 2b HZJF-9041DSchematic diagram of back panel of partial discharge detector

HZJF-9041D Description Of Front Panel And Rear Panel Components

1. Touch screen: display the measured PD waveform, PD, test voltage, test frequency and other parameters;

2. Power switch button: press once to turn on; press again to turn off;

3. Partial discharge input: partial discharge measurement signal input, HZJF-9041D has four channels, the signal can be accessed from any channel;

4. High voltage voltage divider, voltage range: 150V;

5. Grounding bolt: reliable grounding is required during test;

6. RS232 port: through this serial port, the parameters of partial discharge instrument can be uploaded to other equipment;

7. External power supply input;

 

Main Technical Indicators

1. The equivalent capacitance ranges from 6pF to 250 μ F

2. Number of channels: Four channels

3. Measurement range: 0.1pc-10000nc

4. Detection sensitivity and allowable current (see Table 1)

 

 

NO.

Tuning Capacitance Range

Sensitivity (PC)

(Unbalanced Circuit)

Effective Value Of Allowable Current

Unbalanced Circuit

Balance Circuit

1

0~25~100pF

0.02

30mA

0.25A

2

25~100~400pF

0.04

50mA

0.5A

3

100~400~1500pF

0.06

120mA

1A

4

400~1500~6000pF

0.1

0.25A

2A

5

1500~6000~25000pF

0.2

0.5A

4A

6

0.006~0.025~0.1μF

0.3

1A

8A

7

0.025~0.1~0.4μF

0.5

2A

15A

8

0.1~0.4~1.5μF

1

4A

30A

9

0.4~1.5~6.0μF

1.5

8A

60A

10

1.5~6.0~25μF

2.5

15A

120A

11

6.0~25~60μF

5

25A

200A

12

25~60~250μF

10

50A

300A

7R

Resistance

0.5

2A

15A

Table 1. Detection sensitivity and allowable current value of input unit

5. Elliptic scan time base

(1) Frequency 50/60, 100, 150, 200, 400Hz

(2) Rotation: 30 ° as the first gear, and can rotate 120 °

(3) Working method: ellipse sine wave straight line

6. Display unit

10.1 inch capacitive touch screen is used.

7. Amplifier

(1) 3dB low frequency fl:10, 20, 40KHz optional

(2) 3dB high frequency fh:80, 200, 300kHz optional

(3) Gain adjustment, rough adjustment 6, inter gear gain difference 20 ± 1dB, fine adjustment range >20db

(4) Asymmetry of positive and negative pulse response < 1dB

8. Time window

(1) Window width: adjustable, 5 ° ~ 170 ° at 50Hz

(2) Window position: each window can rotate 0 ° ~ 170 °

(3) Two time windows can be opened separately or simultaneously.

9. Peak pulse display

The touch screen displays 1 decimal place (10pc or more), 2 decimal places (less than 10pc), error: ± 3% (in full scale)

10. Test voltage display

(1) Range 150kV

(2) Input impedance: >1m Ω

(3) Display: touch screen display, display 1 decimal place

(4) Error: ± 1%

11. Test frequency display

Error: less than ± 1%

12. Zero standard system

Zero sign is consistent with all ellipse scanning frequencies

13. Structure

(1) Dimensions: 370mm (width), 460mm (depth), 215mm (height)

(2) Weight: about 12.5kg