Troubleshooting and maintenance of insulating oil withstand voltage tester

Jun 04, 2022

Troubleshooting of insulating oil withstand voltage tester:

 

1 The insulating oil withstand voltage tester has no display

 

Approach:

 

1.1 Check whether there is a power supply

 

1.2 Check whether the fuse is blown, one fuse is located in the power socket, and the other is located on the panel.

 

2 After displaying "running", the voltage will not be boosted, nor will it alarm

 

Approach:

 

The fuse on the panel is blown, replace it with a 5A fuse

 

3 When printing, the font is very small, and the paper does not feed

 

Approach:

 

The printing paper roll is too tight, open the printer panel and loosen the printing paper a few times.

 

4 The handwriting is very unclear

 

Approach:

 

The printer ribbon has been used up, replace the printer ribbon

 

Instrument maintenance:

 

1. The oil cup and electrode should be kept clean, and must be protected by filling new transformer oil during the out-of-service period. After the inferior oil is tested, it must be washed with solvent gasoline or carbon tetrachloride and dried before continuing to use.

 

2. After the oil cup and electrode have been used continuously for one month, an inspection should be carried out. Check whether there is any change in the distance between the measuring electrodes. Use a magnifying glass to observe whether there is any darkening on the surface of the electrodes. If this phenomenon occurs, readjust the distance and clean the electrodes with chamois or silk. In case of long-term discontinuation, this work should also be carried out before use.

 

3 If the instrument is not used for a long time, please power on it once a month for 1 hour.

 

4 When the instrument does not work normally, please check whether the fuse has been blown, and the test can be continued after replacing the fuse of the same type.