Turbidity is the degree of obstruction that occurs when suspended matter in water transmits light. The water contains suspended solids and colloids such as soil, dust, fine organic matter, zooplankton and other microorganisms, which can cause turbidity in the water.
Turbidity is one of the sensitive indicators of water, and it is also one of the signs that the water body may be polluted. High water turbidity will affect the survival of aquatic organisms. In general, the measurement of turbidity is mainly used for natural water, drinking water and some industrial water. In sewage treatment, economical and effective coagulants are often selected by measuring turbidity, and the purpose of adjusting the amount of chemicals added at any time to obtain good effluent quality.
Turbidimeter, also known as turbidimeter. It can be used to test the turbidity of water samples in water plants, power plants, industrial and mining enterprises, laboratories and field sites, so it is very important to master the correct method of use.
Precautions for the use of turbidimeters
1. Carefully check the turbidity standard plate. If there is dust or stains, it can be wiped with absorbent cotton and half mixture of ethanol and ether. The cuvette can be cleaned with detergent or detergent, and then rinsed with clean water. The two light-transmitting surfaces Wipe dry. Warm up the instrument for 10 minutes.
2. If the selection key of "Range" needs to be switched during the measurement, except for the selection of different cuvettes, the zero-adjustment step for turbidity-free water must be repeated once.
3. During the measurement, the two optical paths of the cuvette must be free of any dirty spots, and there must be no water stains on the two sides and bottom.
4. When using an old cuvette to hold the sample and add "0" water during measurement, it must be put into "0" water at the same time before the sample is tested.
5. Measure the other position difference (cylinder difference) of the two sample cups, and make corresponding corrections in the subsequent sample turbidity measurement values.
6. If there are few samples to be measured, the same cuvette can be used for zero calibration and sample measurement, so that no matter whether the agency reports the old cuvette, it will not be brought into the cylinder, which is beneficial to the reliability of the test data and avoids unnecessary operation errors.