HZ-9025B Portable Partial Discharge Testing Kit Price
Huazheng HZ-9025B Partial Discharge Test Kit
Introduction
The HZ-9025B partial discharge test kit combines high sensitivity with exceptional adaptability, featuring:
Broad compatibility with diverse test object capacitances
Wide input impedance range (up to 1000 MΩ)
9 configurable frequency bands
Universal application across multiple test objects
Features:
The partial discharge signal is digitally processed and displayed on the LCD screen.
The time base can be freely selected as ellipse, line, or sine mode.
The digital window in the HZ-9025B can be used to avoid noise interference from partial discharge signals.
Single or double windows can be selected at any phase. The ellipse rotation angle is 150°.
The current partial discharge picture can be saved for subsequent pulse analysis.
The current partial discharge result can be saved as an image while retaining the test data.
Primary Technical Specifications
|
No. |
Tuning capacitance range |
Sensitivity (pC) (Un-balance circuit) |
RMS Of max. current |
|
|
(Un-balance circuit) |
(Balance circuit) |
|||
|
1 |
0~25~100pF |
0.02 |
30mA |
0.25A |
|
2 |
25~100~400pF |
0.04 |
50mA |
0.5A |
|
3 |
100~400~1500pF |
0.06 |
120mA |
1A |
|
4 |
400~1500~6000pF |
0.1 |
0.25A |
2A |
|
5 |
1500~6000~25000pF |
0.2 |
0.5A |
4A |
|
6 |
0.006~0.025~0.1μF |
0.3 |
1A |
8A |
|
7 |
0.025~0.1~0.4μF |
0.5 |
2A |
15A |
|
8 |
0.1~0.4~1.5μF |
1 |
4A |
30A |
|
9 |
0.4~1.5~6.0μF |
1.5 |
8A |
60A |
|
10 |
1.5~6.0~25μF |
2.5 |
15A |
120A |
|
11 |
6.0~25~60μF |
5 |
25A |
200A |
|
12 |
25~60~250μF |
10 |
50A |
300A |
|
7R |
Resistor |
0.5 |
2A |
15A |
Working Principle
The partial discharge test operates on the ERA (Electronic Rectangular Analysis) principle.
When partial discharge occurs in the test object Cx under the applied test voltage, the input unit extracts the pulse signal. This signal is then amplified by the front amplifier and filtered through the filter amplifier's frequency band selection. After filtering, the signal is amplified to the required amplitude by the main amplifier. The amplified signal is then sampled by the capturing board and displayed on the screen.
The test voltage value is obtained from the high-voltage divider and supplied to the digital voltmeter.
The working principle of the system is illustrated in the diagram below (Figure 3).

Important Operating Notes
Overload Indication
This device displays partial discharge values on a digital meter with a full graduation designated at 100.0. Values exceeding this indicate overload. When overload occurs, accuracy under the current grade is not guaranteed. The operator should switch the gain to a lower grade.
Noise Suppression
During testing, some noise may occur. For phase-fixed noise, it can be avoided using the window function. Use the left and right window switches and knobs to adjust the window position and width, so as to cover the partial discharge signals while avoiding the noise.



Hot Tags: hz-9025b portable partial discharge testing kit price, China, suppliers, manufacturers, price, for sale, ultrasonic partial discharge detector, pd detection partial discharge measurement, partial discharge detector, partial discharge analyzer, Four Channels Partial Discharge Detector, portable partial discharge test equipment
Send Inquiry