What is the flash and fire point by Cleveland Open Cup apparatus?

Jan 28, 2026

Flash Point and Fire Point (Cleveland Open Cup, COC)

The Cleveland Open Cup (COC) is a standard laboratory apparatus for determining the flash point and fire point of petroleum products and other combustible liquids with relatively high boiling points (typically >79°C / 175°F). It is an open-cup method (ASTM D92, ISO 2592 are the primary international standards), distinct from closed-cup methods (e.g., Pensky-Martens) used for lower-boiling liquids.

 

1. Flash Point (COC)

Definition: The lowest temperature at which the vapor of a liquid, when heated in the Cleveland open cup, mixes with air to form a flammable mixture that ignites briefly (a flash) when exposed to a standardized test flame.

Key Notes:

The flash is momentary (no sustained combustion) because the vapor concentration is just at the lower flammable limit (LFL) at this temperature.

Open-cup flash points are higher than closed-cup flash points for the same liquid (vapors escape to the atmosphere, requiring more heat to build up a flammable mixture).

 

2. Fire Point (COC)

Definition: The lowest temperature at which the vapor of a liquid, heated in the Cleveland open cup, produces enough vapor to ignite and burn steadily for at least 5 seconds when exposed to the test flame.

Key Notes:

The fire point is always higher than the flash point (typically 5–40°C higher) because it requires a higher vapor concentration to sustain combustion.

This parameter indicates the temperature at which a liquid can release flammable vapors at a rate sufficient for continuous burning-critical for fire safety and handling guidelines.

 

Core Test Procedure (COC Apparatus)

A sample is placed in the cup-shaped metal vessel of the COC apparatus (standard volume, calibrated dimensions).

The sample is heated at a controlled rate (2–3°C/min for flash point; increased after flash point for fire point).

A standardized test flame (small, fixed size) is passed over the surface of the liquid at regular temperature intervals (or continuously for automated units).

Flash point: Record the temperature when a flash first occurs.

Fire point: Continue heating and record the temperature when the vapor burns steadily for ≥5 seconds.

 

Typical Applications

The COC method is used for:

Lubricating oils, hydraulic fluids, fuel oils, asphalt, waxes, and heavy petroleum distillates.

Liquids with flash points above 79°C (175°F) (ASTM D92 covers flash points from 79°C to 400°C+).

Fire safety classification, product labeling, and compliance with occupational/environmental regulations.

 

Key Standards

ASTM D92: Standard Test Method for Flash and Fire Points by Cleveland Open Cup.

ISO 2592: Petroleum products-Determination of flash and fire points-Cleveland open cup method.

IP 36: UK standard equivalent to ASTM D92/ISO 2592.

 

Example Values (COC)

Product Flash Point (COC) Fire Point (COC)
Heavy fuel oil (No. 6) ~150–200°C ~160–210°C
Motor oil (10W-40) ~200–250°C ~210–260°C
Mineral hydraulic oil ~180–220°C ~190–230°C

 

Critical Distinction: Open vs. Closed Cup

Aspect Cleveland Open Cup (COC) Closed Cup (e.g., Pensky-Martens)
Flash point value Higher Lower
Liquid type High-boiling, viscous Low-to-medium boiling
Vapor loss Significant (open system) Minimal (sealed system)
Primary standard ASTM D92/ISO 2592 ASTM D93/ISO 2719