The solution to the blockage of the injection needle of the gas chromatograph

Apr 17, 2022

In response to this situation, you can start from the following aspects:

1) Reinstall the autosampler plunger.

2) The wear of the plunger of the GC gas chromatograph autosampler includes both the number of full-scale cycles and the remaining sample remaining in the syringe. Counterintuitively, full-scale injections are more likely to wear out the plunger than half-scale injections. If the sample contains non-volatile sample residues that can deposit in the injector cavity.

3) Fill the injector column with pure solvent free of particles. Let stand for a few minutes for the solvent to dissolve the soluble material. This material is preferably squeezed through a needle into a heated vacuum injector cleaning device. Repeat this operation.

4) If the plunger is stuck or clamped in a vertical position of the injector, only a drop of solvent around the plunger can be injected to the top of the injector. These solvents are drawn into the space between the injector lumen and the piston by siphoning. Such leaks can last for hours or more.

5) When GC gas chromatograph usually chooses the type of solvent, first choose a non-polar solvent (such as hexane), and then choose a strong polar solvent. Move the plunger and wipe gently with a clean paper towel. (Do not pull the plunger of the injector out of the needle.) Before replacing the plunger, insert the needle into the heated injection port and allow the heated carrier gas to dry the needle and cavity (prevent air from entering the injector).