Cleaning method of oil cup and stirring paddle of transformer oil tester

May 14, 2022

1. Cleaning method of oil cup

(1) Wipe the electrode surface and electrode rod repeatedly with clean silk;

(2) Adjust the electrode gap with standard gauges;

(3) Clean it with petroleum ether or ethanol for 3 to 4 times, then dry it with a hair dryer. It can be cleaned 2 or 3 times after putting in the tested oil sample.

2. Cleaning method of stirring paddle

(1) Wipe the stirring paddle repeatedly with a clean silk cloth until there are no fine particles on the surface, and avoid touching the surface of the stirring paddle with your hands;

(2) Clamp the stirring paddle with tweezers, immerse it in absolute ethanol, rinse repeatedly, and blow dry with a hair dryer;

(3) Clamp the stirring paddle with tweezers and immerse the oil sample to be tested for repeated scrubbing