Transformer Oil Auto Open Cup Flash Point Testing Instrument
Product Introduction
Our company's production lines are all introduced from abroad to ensure high standard and production quality of the insulation resistance tester meter, oil immersed test transformer, power quality and energy analyzer. On the basis of maximizing the benefits of the company and meeting the needs of customers, we show our personal creativity. We adopt an optimized product structure design and manufacturing process. If you have the honor to do business with our company.
HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus
Product Overview
The Automatic Cleveland Open Cup Flash Point Apparatus are suitable for the determination of petroleum products according to the standard petroleum products flash point and ignition point determination method (Cleveland open cup method) except for fuel oil and petroleum products with open flash point lower than 79℃.
The instrument is designed as a table-top structure, with stainless steel working table and stainless steel heating furnace shell, which is beautiful and elegant. The Automatic Cleveland Open Cup Flash Point Apparatus are equipped with flame extinguishing cover, which meets the requirements of the ignition point test.
Features
We believe that our Automatic Cleveland Open Cup Flash Point Apparatus can compete with any other similar products in terms of quality performance as well as price. It adopts a special heating furnace form to ensure the safety of the test. And its heating power is continuously adjustable to meet the requirements of the test. In addition, the instrument is complete, the user only needs to match the gas or other civil gas can be tested, it is very simple and convenient to use.
HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus

Technical Parameters
| Temperature Measurement | |
Temperature measuring range | -59.9°C~399.9°C |
Temperature Control Range | Room temperature~400°C |
Repeatability | ≤150℃ 3℃, >150℃ 5℃ |
Resolution | 0.1°C |
Precision | 0.5% |
Temperature Sensor | Platinum Resistor(PT100) |
Flash Fire Sensor | Ion detection ring |
Ambient Temperature | 0 ~ 40°C |
Relative Humidity | ≤ 85% RH |
Power Supply | AC220V ± 20%; 50HZ ± 10% |
Power Consumption | Approx 400W |
Overall Dimension | 440X280X290(mm) |
Instrument Weight | 14kg |
Failure And Treatment
NO. | Failure | Treatment |
1 | Turn on the power, the LCD screen no display | Check power supply is plugged in, open the instrument measured the board examines the plug is not loose. |
2 | Large repeatability error for sample testing | Remove the lifting arm top cover, and check |
3 | No heating | Electric wire broken |
4 | Lift arm rise or fall failure | Limit opt coupler damage |








We are dedicated to consistently providing our customers in every industry with outstanding service and competitive pricing on our top-quality Transformer Oil Auto Open Cup Flash Point Testing Instrument. Believe that we will provide you with excellent products, services and technical knowledge, and help your business thrive. We actively undertake the social responsibility of enterprises and undertake our obligations and responsibilities in the direction of sustainable development of industrial society. It is a pleasure to have friends from afar. We welcome colleagues from all over the world to further strengthen exchanges and cooperation with us.
Hot Tags: HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus, China, suppliers, manufacturers, price, for sale, Automatic Cleveland Open Cup Flash Point Tester, flashpoint tester cleveland, Open Cup Flash Point Analyzer, opencup flash point analyser, Cleveland Open-Cup (COC) measure, HZKS-3, hzks3, open cup flash fire point tester, Circuit Breaker Analyser, ground down lead conduction tester, protection relay test procedure, transformer capacity noload load test, surge arrester testing, transformer oil purification equipment
[[ProKeywords]]SPEC DATA