Biobase Open-Cup Flash Point Tester
Product Introduction
Our company has a well-organized, efficient, reliable team. We are able to design and develop high-quality ct pt analyser, closed cup flash point analyzer, test transformer 100kv, we would greatly appreciate the chance to convince you of our quality, ability and potential. We continue to improve corporate governance, optimize mechanisms and processes, and improve service levels and operational efficiency. We respect the individuality of our employees, invest in their competitive ability and happy work, and give reasonable returns for their contributions.
HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus
Product Overview
The Automatic Cleveland Open Cup Flash Point Apparatus are suitable for the determination of petroleum products according to the standard petroleum products flash point and ignition point determination method (Cleveland open cup method) except for fuel oil and petroleum products with open flash point lower than 79℃.
The instrument is designed as a table-top structure, with stainless steel working table and stainless steel heating furnace shell, which is beautiful and elegant. The Automatic Cleveland Open Cup Flash Point Apparatus are equipped with flame extinguishing cover, which meets the requirements of the ignition point test.
Features
We believe that our Automatic Cleveland Open Cup Flash Point Apparatus can compete with any other similar products in terms of quality performance as well as price. It adopts a special heating furnace form to ensure the safety of the test. And its heating power is continuously adjustable to meet the requirements of the test. In addition, the instrument is complete, the user only needs to match the gas or other civil gas can be tested, it is very simple and convenient to use.
HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus

Technical Parameters
| Temperature Measurement | |
Temperature measuring range | -59.9°C~399.9°C |
Temperature Control Range | Room temperature~400°C |
Repeatability | ≤150℃ 3℃, >150℃ 5℃ |
Resolution | 0.1°C |
Precision | 0.5% |
Temperature Sensor | Platinum Resistor(PT100) |
Flash Fire Sensor | Ion detection ring |
Ambient Temperature | 0 ~ 40°C |
Relative Humidity | ≤ 85% RH |
Power Supply | AC220V ± 20%; 50HZ ± 10% |
Power Consumption | Approx 400W |
Overall Dimension | 440X280X290(mm) |
Instrument Weight | 14kg |
Failure And Treatment
NO. | Failure | Treatment |
1 | Turn on the power, the LCD screen no display | Check power supply is plugged in, open the instrument measured the board examines the plug is not loose. |
2 | Large repeatability error for sample testing | Remove the lifting arm top cover, and check |
3 | No heating | Electric wire broken |
4 | Lift arm rise or fall failure | Limit opt coupler damage |








We are committed to creating top-notch Biobase Open-Cup Flash Point Tester, creating maximum benefits, and making every effort to build an international first-class large-scale industrial group. Our company always adheres to the business philosophy of 'professional products and professional services', regards product quality as the life of the enterprise, and strives to provide reasonable prices and good services for customers. Our goal is to create a national brand, and at the same time to catch up with the world's scientific and technological level.
Hot Tags: HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus, China, suppliers, manufacturers, price, for sale, Automatic Cleveland Open Cup Flash Point Tester, flashpoint tester cleveland, Open Cup Flash Point Analyzer, opencup flash point analyser, Cleveland Open-Cup (COC) measure, HZKS-3, hzks3, open cup flash fire point tester, oil breakdown voltage tester 3 cup, test transformer 100kv, Ampere Second Characteristic Tester, on load tap switch tester, hv resonant test, Contact resistance test equipment