Automatic Open Cup Flash Point Testing Meter with Self-Checking Function
Product Introduction
We are an enterprise specializing in high voltage circuit breakers analyzer, kinematic viscosity bath, load loss test of transformer. We can select the most appropriate products for our customers in terms of price, brand and performance, and provide reliable, systematic and perfect after-sales service. As the main creators of our important human capital and value, our employees are the inexhaustible power and source of our sustainable development and competition. The direct purpose of the establishment and existence of the company is to maximize profits and further achieve long-term development and growth. Corporate culture is the soul of an enterprise, thus we strive to build a corporate culture with the spirit of the times and its own characteristics.
HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus
Product Overview
The Automatic Cleveland Open Cup Flash Point Apparatus are suitable for the determination of petroleum products according to the standard petroleum products flash point and ignition point determination method (Cleveland open cup method) except for fuel oil and petroleum products with open flash point lower than 79℃.
The instrument is designed as a table-top structure, with stainless steel working table and stainless steel heating furnace shell, which is beautiful and elegant. The Automatic Cleveland Open Cup Flash Point Apparatus are equipped with flame extinguishing cover, which meets the requirements of the ignition point test.
Features
We believe that our Automatic Cleveland Open Cup Flash Point Apparatus can compete with any other similar products in terms of quality performance as well as price. It adopts a special heating furnace form to ensure the safety of the test. And its heating power is continuously adjustable to meet the requirements of the test. In addition, the instrument is complete, the user only needs to match the gas or other civil gas can be tested, it is very simple and convenient to use.
HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus

Technical Parameters
| Temperature Measurement | |
Temperature measuring range | -59.9°C~399.9°C |
Temperature Control Range | Room temperature~400°C |
Repeatability | ≤150℃ 3℃, >150℃ 5℃ |
Resolution | 0.1°C |
Precision | 0.5% |
Temperature Sensor | Platinum Resistor(PT100) |
Flash Fire Sensor | Ion detection ring |
Ambient Temperature | 0 ~ 40°C |
Relative Humidity | ≤ 85% RH |
Power Supply | AC220V ± 20%; 50HZ ± 10% |
Power Consumption | Approx 400W |
Overall Dimension | 440X280X290(mm) |
Instrument Weight | 14kg |
Failure And Treatment
NO. | Failure | Treatment |
1 | Turn on the power, the LCD screen no display | Check power supply is plugged in, open the instrument measured the board examines the plug is not loose. |
2 | Large repeatability error for sample testing | Remove the lifting arm top cover, and check |
3 | No heating | Electric wire broken |
4 | Lift arm rise or fall failure | Limit opt coupler damage |








We are committed to providing the most valuable Automatic Open Cup Flash Point Testing Meter with Self-Checking Function and service solutions for all industries. Our company regards 'reasonable prices, efficient production time and good after-sales service' as our tenet. We continue to upgrade our technology to promote the progress of environmental protection in traditional manufacturing industries, and hope to make a great contribution to the sustainable development of the whole industry.
Hot Tags: HZKS-3 Automatic Cleveland Open Cup Flash Point Apparatus, China, suppliers, manufacturers, price, for sale, Automatic Cleveland Open Cup Flash Point Tester, flashpoint tester cleveland, Open Cup Flash Point Analyzer, opencup flash point analyser, Cleveland Open-Cup (COC) measure, HZKS-3, hzks3, open cup flash fire point tester, vlf hypoctic tester, Transformer Oil Purifier, transformer oil dissolved gas analyzer, huazheng hz3340d, sf6 test, partial discharge measurement