HZCT-100C High Accuracy 0.05% CT PT Analyzer
Panel
The instrument panel structure is as shown in right figure.
• Red S1, black S2 terminal: test power output
• Yellow M1, black M2 terminal: measure output voltage
• Green P1, black P2 terminal: measure inductive voltage

Technical Parameter
|
|
Transformer Tester |
|
|
Usage |
CT & VT |
|
|
Output |
0~180Vrms,12Arms,18A(peak value) |
|
|
CT Ratio |
Range |
1~40000 |
|
accuracy |
±0.05% |
|
|
PT Ratio |
Range |
1~40000 |
|
accuracy |
±0.2% |
|
|
Phase |
Range |
±5min |
|
accuracy |
0.5min |
|
|
DC resistance |
Range |
0~300Ω |
|
accuracy |
2%±2mΩ |
|
|
Burden |
Range |
0~300VA |
|
accuracy |
2%±0.2VA |
|
|
Power supply |
AC220V±10%,50/60Hz |
|
|
Environmental Conditions |
Operating temperature :-10οC ~50οC,Humidity :≤90% |
|
|
Weight and Dimensions |
Dimensions:340 mm×300 mm×150mm,Weight<9kg |
|