◆Analysis of dissolved gases in insulating oil
Content determination of dissolved gases in insulating oil by gas chromatography is set, the power supply enterprise to judge the oil filled power equipment in the operation of the existence of latent heat, discharge fault, in order to ensure the effective operation of the power grid safety, It is also one of the necessary means of testing equipment for oil-filled electrical equipment manufacturers.HZGC-1212A transformer oil chromatographic analysis system using three detector process standard recommend, a sampling can complete component of dissolved gas in insulation oil (including hydrogen, oxygen, methane, ethylene, ethane, acetylene, carbon monoxide and carbon dioxide) full content analysis.
◆Determination of gas content in insulating oil
Insulating oil gas content is an important index of oiliness supervision. At present, according to the DL/T450-1991 method of making carbon dioxide elution method is suitable for determining not only acid gas containing oil, according to the DL/T423-91 method for vacuum pressure difference method for vacuum instrument is not easy and the existence of the limitations.HZGC-1212A transformer oil spectrum analysis of process design of the system fully in line with the People's Republic of China electric power industry standard DL/T 703-1999< gas content in insulating oil by gas chromatography method in the chromatographic process design. The machine is equipped with the provisions. The thermal conductivity detector with high sensitivity and hydrogen flame ionization detector, and a nickel catalyst converter, can be dissolved in transformer oil nine gas components: hydrogen, oxygen, nitrogen, carbon monoxide, carbon dioxide, methane, ethane, ethylene, acetylene measuring. Its performance to meet the requirements of gas insulated DL/T 703-1999< chromatograph oil and gas content in gas chromatography ".
✿Instrument configuration
Name | Model | Number | Remark |
Gas chromatograph | HZGC-1212A | 1platform | Standard accessories |
Detector | FID | 2individual | Standard accessories |
TCD | 1individual | ||
Conversion device | 1The cover | Standard accessories | |
Chromatographic column | 1The cover | Standard accessories | |
Chromatography workstation | Special power system | 1The cover | Standard accessories |
Oscillator instrument | 1platform | Buy locall or we can provide | |
Standard gas | 1bottle | Buy locally | |
Gas source | Argon, hydrogen, air | 1The cover | Buy locally or we can provide the gas generator |
Computer, printer | 1The cover | Buy locally |
✿Instrument performance
The first sampling, the sample volume is 1 mL, the minimum detection concentration of transformer oil(ul/L):
Component | H2 | O2 | N2 | CO | CO2 | CH4 | C2H4 | C2H6 | C2H2 |
Minimum detectable concentration(ul/L) | 2 | 10 | 20 | 1 | 5 | 0.1 | 0.1 | 0.1 | 0.1 |